TECHNICAL DATASHEET Medical AC-DC Adapter MDS-030AAC Series / MDS-030AAC 030AAC Highlights & Features MDS-030AAC05, 12, 15, 24 Meet Efficiency Level VI MDS-030AAC07 Meet Efficiency Level V Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1 Compliant with IEC 60601-1-2 4th Ed. Requirements Meets Limited Power Source (LPS) requirements Low touch current. Suitable for type BF applications Detachable AC plug with multiple country options IP22 ingress protection rating 2 MOPP isolation 500K hours MTBF Safety Standards CB Certified for worldwide use Model Number: MDS-030AAC Unit Weight: 150 grams Dimensions (W L H): 53.5 88.0 27.5 mm 2.11 3.46 1.08 inch General Description The MDS series of external power supply comes with universal AC input at 85Vac to 264Vac. Other features include low touch current, risk management report available and the electric shock protection comply with 2 MOPP. The MDS series is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN 55032 for Information Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used. The MDS series come with both medical and ITE safety approvals including UL/cUL and IEC-60950-1 CB certification as a Limited Power Source (LPS), and CCC approval. Designs are fully compliant with RoHS Directive 2011/65/EU for environmental protection. Model Information Medical AC-DC Adapter Model Number Input Voltage Range Efficiency Level Rated Output Voltage Rated Output Current MDS-030AAC05 85-264Vac Level VI 5Vdc 3A MDS-030AAC07 Level V 7Vdc 3A MDS-030AAC12 Level VI 12Vdc 2A MDS-030AAC15 Level VI 15Vdc 2A MDS-030AAC24 Level VI 24Vdc 1.25A Model Numbering 030 MDS AAC 1 2 Delta Medical Max wattage in the Family Code Output Voltage (Single Revision Code Revision Code Power Supply product series Output) DC plug type and output cable length Country duck-head type - 05 for 5V - 07 for 7V A 2.1 x 5.5 x10mm DC plug A China B United States - 12 for 12V B 2.1 x 5.5 x12mm DC plug C European D United Kingdom - 15 for 15V Both A & B are 1200mm cable length - 24 for 24V E Australia GKorea H India, J Argentina K Brazil MSouth Africa * MDS-030AAC15 BB is not available. All parameters are specified at 25C ambient unless otherwise indicated. 1 www.DeltaPSU.com (July 2019, Rev.10) TECHNICAL DATASHEET Medical AC-DC Adapter MDS-030AAC Series / MDS-030AAC Specifications Model Number MDS-030AAC05 MDS-030AAC07 MDS-030AAC12 MDS-030AAC15 MDS-030AAC24 Input Ratings / Characteristics Nominal Input Voltage 100-240Vac Input Voltage Range 85-264Vac Nominal Input Frequency 50-60Hz Input Frequency Range 47-63Hz Input Current (max) 115Vac 0.5A 0.6A 0.8A 0.8A 0.8A 230Vac 0.3A 0.4A 0.6A 0.6A 0.6A Average Efficiency (min) 115Vac 81.4% 81.4% 86.21% 87.0% 87.0% 230Vac Standby Power (max) 115Vac 0.1W 0.1W 0.1W 0.1W 0.1W 230Vac Inrush Current (typ.) 60A 115Vac and 100A 230Vac 1) 2) Touch Current (max) 0.1mA 264Vac NC , 0.3mA 264Vac SFC 1) 2) Output-PE (protective earth) 0.1mA 264Vac NC , 0.5mA 264Vac SFC leakage current for Type BF application (max) 1) NC: Normal condition 2) SFC: Single fault condition Model Number MDS-030AAC05 MDS-030AAC07 MDS-030AAC12 MDS-030AAC15 MDS-030AAC24 Output Ratings / Characteristics Nominal Output Voltage 5Vdc 7Vdc 12Vdc 15Vdc 24Vdc Output Total Regulation 6% Output Current 3A 3A 2A 2A 1.25A Output Power 15W 21W 24W 30W 30W Line Regulation 1% Load Regulation 5% Ripple & Noise (max) 100mVpk-pk 100mVpk-pk 150mVpk-pk 200mVpk-pk 200mVpk-pk Rated load Rated load Rated load Rated load Rated load Start-up Time (max) 3000ms 115Vac Hold-up Time (min) 10ms 115Vac Dynamic Response (Overshoot & 10% 50-100% load Undershoot O/P Voltage) Capacitive load (max) 4700uF *Periodic and Random Deviation. Ripple & Noise measurement circuit with 20MHz BW PSU Load 0.47uF Ceramic Scope 5cm 10uF Electrolytic Test Fixture All parameters are specified at 25C ambient unless otherwise indicated. 2 www.DeltaPSU.com (July 2019, Rev.10)