TECHNICAL DATASHEET MDS Medical AC-DC Wall Mount Adapter 5Volt, 5Watt / MDS-005AAS05 B 005AAS05 B Highlights & Features Meet Efficiency Level VI Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 62368-1 Compliant with IEC 60601-1-2 4th Ed. Requirements IP22 Ingress Protection Rating 1 Million Hours MTBF Safety Standards CB Certified for worldwide use Model Number: MDS-005AAS05 B Unit Weight: 100 grams (3.6 ounces) Dimensions (W x L x H): 39.5 x 56.5 x 28.0 mm (1.56 x 2.23 x 1.1 inch) General Description The MDS series of external power supply come with universal AC input at 90Vac to 264Vac. Other features include low touch current, risk management report available and the electric shock protection comply with 2 x MOPP. The MDS series is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN 55032 for Information Technology Equipment (ITE) radio-frequency equipment. The MDS series of external power supply come with both medical and ITE safety approvals including CE/CB certification and are fully compliant with RoHS Directive for environmental protection. Model Information Medical AC-DC Charger Model Number Input Voltage Range Output Voltage Output Current MDS-005AAS05 B 90-264Vac 5Vdc 1A Model Numbering - 05 005 AAS MDS B Family Code Output Voltage Output Power* Delta Medical Control Code Single Output: BV : US Type with Barrel 005 5W Power Supply 05 for 5V Type connector BU : US Type with Micro USB connector * Maybe lower at some voltage. All parameters are specified at 25C ambient unless otherwise noted. 1 www.DeltaPSU.com (September 2020, Rev. 05) TECHNICAL DATASHEET MDS Medical AC-DC Wall Mount Adapter 5Volt, 5Watt / MDS-005AAS05 B Specifications Input Ratings / Characteristics Nominal Input Voltage 100-240Vac Input Voltage Range 90-264Vac Nominal Input Frequency 50-60Hz Input Frequency Range 47-63Hz Input Current (max) 0.2A 115Vac 0.1A 230Vac Efficiency (typ.) 71%, Reference Fig.1 Average Efficiency (min) 73.63% 115Vac and 230Vac Standby Power (max) 0.1W 115Vac and 230Vac Inrush Current (typ.) No component damage 1) 2) Touch Current (max) 0.1mA 264Vac NC , 0.3mA 264Vac SFC 1) NC: normal condition 2) SFC: single fault condition 100 95 90 85 80 115Vac Efficiency 75 230Vac Efficiency 70 65 60 55 50 25% 50% 75% 100% Output load Fig.1 Efficiency versus output load Output Ratings / Characteristics Nominal Output Voltage 5Vdc Total regulation 10% Output Current 1A Output Power 5W Line Regulation (max) 1% Load Regulation (max) 9% Ripple & Noise (typ.) 88mV pk-pk Full load, Reference Fig. 3 & 4 Start-up Time(max) 3000ms 115Vac Hold-up Time(min) 5ms 115Vac Dynamic Response (Overshoot & Undershoot O/P Voltage) 10% 50-100% load (Slew Rate: 0.1A/S, 50% duty cycle 1KHz) *Periodic and Random Deviation All parameters are specified at 25C ambient unless otherwise noted. 2 www.DeltaPSU.com (September 2020, Rev. 05) E fficiency (% )